Abstract of "Muscular Bio-Stimulator" Project
This is a small, portable set, designed for those aiming at look improvement. The Bio-Stimulator provides muscles' stimulation and invigoration but, mainly, it's an aid in removing cellulitis.
Tape the electrodes to the skin at both ends of the chosen muscle and rotate P1 knob slowly until a light itch sensation is perceived. Each session should last about 30 - 40 minutes.The use of this device is forbidden to Pace-Maker bearers and pregnant women.
Don't place the electrodes on cuts, wounds, injuries or varices.
Obviously we can't claim or prove any therapeutic effectiveness for this device.
IC1 generates 150΅Sec. pulses at about 80Hz frequency. Q1 acts as a buffer and Q2 inverts the pulses' polarity and drives the Transformer. Output pulses' amplitude is set by P1 and approximately displayed by LED D1 brightness. D2 protects Q2 against high voltage peaks generated by T1 inductance during switching.
Features of Sleep Inducer
• T1 is a small mains transformer 220 to 12V @ 100 or 150mA. It must be reverse connected i.e.: the 12V secondary winding to Q2 Collector and ground, and the 220V primary winding to output electrodes.
• Output voltage is about 60V positive and 150V negative but output current is so small that there is no electric-shock danger.
• In any case P1 should be operated by the "patient", starting with the knob fully counter-clockwise, then rotating it slowly clockwise until the LED starts to illuminate. Stop rotating the knob when a light itch sensation is perceived.
• Best knob position is usually near the center of its range.
• In some cases a greater pulse duration can be more effective in cellulitis treatment. Try changing R2 to 5K6 or 10K maximum: stronger pulses will be easily perceived and the LED will shine more brightly.
• Electrodes can be obtained by small metal plates connected to the circuit's output via usual electric wire and can be taped to the skin. In some cases, moistening them with little water has proven useful.
• SW1 should be ganged to P1 to avoid abrupt voltage peaks on the "patient's" body at switch-on, but a stand alone SPST switch works quite well, provided you remember to set P1 knob fully counter-clockwise at switch-on.
• Current drawing of this circuit is about 1mA @ 3V DC.
• Some commercial sets have four, six or eight output electrodes. To obtain this you can retain the part of the circuit comprising IC1, R1, R2, C1, C2, SW1 and B1. Other parts in the diagram (i.e. P1, R3, R4, D1, D2, Q2 & T1) can be doubled, trebled or quadrupled. Added potentiometers and R3 series resistors must be wired in parallel and all connected from Emitter of Q1 to positive supply.
• Commercial sets have frequently a built-in 30 minutes timer. For this purpose you can use the Timed Beeper the Bedside Lamp Timer or the Jogging Timer circuits available in this Website, adjusting the timing components to suit your needs.
Circuit Diagram
Electronic Components Required
P1______________4K7 Linear Potentiometer
R1____________180K 1/4W Resistor
R2______________1K8 1/4W Resistor (see Notes)
R3______________2K2 1/4W Resistor
R4____________100R 1/4W Resistor
C1____________100nF 63V Polyester Capacitor
C2____________100µF 25V Electrolytic Capacitor
D1______________LED Red 5mm.
D2___________1N4007 1000V 1A Diode
Q1,Q2_________BC327 45V 800mA PNP Transistors
IC1____________7555 or TS555CN CMos Timer IC
T1_____________220V Primary, 12V Secondary 1.2VA Mains Transformer
SW1____________SPST Switch (Ganged with P1)
B1_____________3V Battery (two 1.5V AA or AAA cells in series etc.)
Construction and Working
IC1 generates 150µSec. pulses at about 80Hz frequency. Q1 acts as a buffer and Q2 inverts the polarity of the pulses and drives the Transformer. The amplitude of the output pulses is set by P1 and approximately displayed by the brightness of LED D1. D2 protects Q2 against high voltage peaks generated by T1 inductance during switching.
T1 is a small mains transformer 220 to 12V @ 100 or 150mA. It must be reverse connected i.e. the 12V secondary winding across Q2 Collector and negative ground, and the 220Vprimary winding to output electrodes.
Output voltage is about 60V positive and 150V negative but output current is so small that there is no electric-shock danger.
In any case P1 should be operated by the “patient”, starting with the knob fully counter-clockwise, then rotating it slowly clockwise until the LED starts to illuminate. Stop rotating the knob when a light itch sensation is perceived.
Best knob position is usually near the center of its range.
In some cases a greater pulse duration can be more effective in cellulite treatment. Try changing R2 to 5K6 or 10K maximum: stronger pulses will be easily perceived and the LED will shine more brightly.
Electrodes can be obtained by small metal plates connected to the output of the circuit via usual electric wire and can be taped to the skin. In some cases, moistening them with little water has proven useful.
SW1 should be ganged to P1 to avoid abrupt voltage peaks on the “patient’s” body at switch-on, but a stand alone SPST switch will work quite well, provided you remember to set P1 knob fully counter-clockwise at switch-on.
Current drawing of this circuit is about 1mA @ 3V DC .
Some commercial sets have four, six or eight output electrodes.
To obtain this you can retain the part of the circuit comprising IC1, R1, R2, C1, C2, SW1 and B1.
Other parts in the diagram (i.e. P1, R3, R4, D1, D2, Q2 & T1) can be doubled, trebled or quadrupled.
Added potentiometers and R3 series resistors must be wired in parallel and all connected across Emitter of Q1 and positive supply.
Commercial sets have frequently a built-in 30 minutes timer. For this purpose you can use the Timed Beeper.
Highlights of the Project
This project is useful to relieve headache and muscular pain and revive frozen muscles that impair movement. Though it provides muscles stimulation and invigoration, it’s mainly an aid in removing cellulitis.
Related Projects : Zigbee Based Patient Analysing System with WAP , Application of Stem Cell Therapy in Spinal Cord Injuries , Advanced Patients Monitoring Through CCTV Camera , 3-D Pharmacophore Modelling , Wearable Object Detection System for the Blind , Low Cost Solar ECG with Bluetooth Transmitter , Microphone Array System for Automatic Fall Detection, Real Time Study of Urine Flow Analysis, Effect of Mobile Phone Emissivity on Bacterial Cell Cultures
<<